How does a metabolic cart work: measuring oxygen use breath by breath

A metabolic cart measures how much air you breathe out and how much oxygen and carbon dioxide it contains. Air in contains about 20.9% oxygen, so the oxygen missing from exhaled air shows what the body used. Multiply that by the volume breathed and you get VO2, in litres per minute.

· Accredited Exercise and Sports Scientist, Founder of EvoQuest
Updated · 4 min read
Athlete wearing a mask for gas analysis on a metabolic cart

Key takeaways

  • VO2 is the volume of air breathed times the drop in oxygen fraction from inhaled to exhaled air.
  • Carbon dioxide production divided by oxygen use gives the respiratory exchange ratio (RER).
  • Flow sensors and gas analysers need daily calibration with known volumes and known gas mixtures.
  • Breath-by-breath, mixing-chamber and Douglas bag methods give slightly different results.
  • Mask leaks, humidity and sensor delay are common sources of error.
On this page
  1. The core idea: what goes in and what comes out
  2. The hardware: flow, gas and software
  3. Three ways to collect the air
  4. Calibration and the sources of error
  5. What validation studies found
  6. What this means for athletes
  7. FAQs

The core idea: what goes in and what comes out

Your muscles burn fuel using oxygen and produce carbon dioxide. A metabolic cart never looks at the muscles. It compares the air you breathe in with the air you breathe out. Room air is about 20.93% oxygen and 0.04% carbon dioxide. Exhaled air at rest might be around 16 to 17% oxygen and 4% carbon dioxide, and the numbers change with exercise. The plain-English introduction is in what is a metabolic cart.

A back-of-envelope example: someone breathes out 7 litres of air per minute at rest, and the oxygen fraction has fallen by about 4.4 percentage points (20.93% to 16.5%). Oxygen used is roughly 7 x 0.044 = 0.31 litres per minute. Real carts apply a correction for the fact that inhaled and exhaled volumes differ slightly (the Haldane transformation), because the body takes in a little more oxygen than it gives back carbon dioxide, so the sums are a little more involved than shown here. VO2 max is simply the highest VO2 reached in a test, covered in what is VO2 max.

The hardware: flow, gas and software

The flow sensor sits in the mask or mouthpiece and measures how much air passes with each breath. Common types are the turbine (a small spinning vane), the pneumotach (which measures a pressure drop across a screen) and ultrasonic sensors. The gas analysers sample a small stream of exhaled air through a thin tube. Oxygen analysers often use paramagnetic or electrochemical sensors, and carbon dioxide analysers often use infrared absorption. Exact designs vary by manufacturer.

Software then converts the volume to standard conditions. Air breathed out is warm and humid, and the same number of molecules take up more space when warm. The cart records temperature, pressure and humidity, and reports volumes at STPD (standard temperature and pressure, dry) so that the results do not depend on the weather in the lab. In a hot, humid room in Tamil Nadu, these corrections matter more than in a cool one.

Carbon dioxide production (VCO2) is calculated in the same way as VO2. The ratio VCO2 divided by VO2 is the respiratory exchange ratio. An RER of about 0.7 indicates mostly fat is burned, about 1.0 indicates mostly carbohydrate, and values above 1.0 in a hard test arise partly because the body is also blowing off extra carbon dioxide.

Three ways to collect the air



Time alignment is a hidden challenge in breath-by-breath systems. The gas sample reaches the analyser a fraction of a second after the flow is measured, and each analyser has its own response time. Software must line the two signals up correctly or the calculated VO2 for each breath will be off. Because single breaths are noisy, results are averaged over 15 to 30 seconds before a VO2 value is picked.

Calibration and the sources of error

Every test day starts with calibration. The flow sensor is checked by pumping a syringe of known volume, commonly 3 litres, through it at several speeds. The gas analysers are calibrated with a bottle of gas with a known oxygen and carbon dioxide mix and with room air. If the sensors are off, every number is off by the same percentage. Common error sources include:

  • A mask that leaks, so some breath escapes the flow sensor and VO2 reads low.
  • Water vapour condensing in the sampling line, which blocks or confuses the analyser.
  • Calibration gas that is out of date or a drifting sensor.
  • A mismatch between the flow and gas timing.
  • Hot, humid air that changes sensor behaviour between calibration and test.

Because the reference result for decades was the Douglas bag, published validation work compares carts to it.

What validation studies found

One study validating a portable system against the Douglas bag and a stationary cart found that its mixing-chamber and breath-by-breath modes were within small VO2 margins of the Douglas bag. Compared with the stationary cart, however, the breath-by-breath mode ran about 6.6% below at the highest cycling load, while the mixing-chamber mode ran about 5.8% above. A comparison of the ParvoMedics TrueOne 2400 and MedGraphics VO2000 systems in the European Journal of Applied Physiology looked at accuracy and reliability, and a validation of the Aerosport KB1-C portable system against the Douglas bag appeared in the International Journal of Sports Medicine.

The practical reading is that systems differ, so you should repeat tests on the same cart with the same protocol, and treat a gap of a few percent between machines as ordinary.

What this means for athletes

A metabolic cart is a specialised gas analysis device. It is valuable where the exact oxygen cost of work matters, such as in endurance sport. Many field and school athletes get more useful, cheaper information from lactate testing and heart rate. The athlete's view is in metabolic cart for athletes, and the VO2 max context is in VO2 max testing for athletes.

EvoQuest's three-hour assessment in Ramanathapuram, Coimbatore, measures aerobic fitness through lactate testing and heart rate monitoring, which can set individual training zones. Not every athlete needs gas analysis, and a coach can explain which tests suit your sport.

If you want to know whether a test like this suits your sport or your child, start with a free 20-minute call on the call page or WhatsApp +91 93440 48834.

Frequently asked questions

How does a metabolic cart measure VO2?

It measures the air you breathe out and the fraction of oxygen and carbon dioxide in it. Because inhaled air is about 20.9% oxygen, the oxygen missing from exhaled air is what the body used. Multiplying by the volume breathed gives VO2 in litres per minute.

What is the respiratory exchange ratio?

The respiratory exchange ratio (RER) is carbon dioxide produced divided by oxygen used. A value near 0.7 suggests mostly fat is the fuel, and near 1.0 suggests mostly carbohydrate. Above 1.0 in hard exercise partly reflects extra carbon dioxide being blown off.

Why must a metabolic cart be calibrated every day?

The flow sensor and gas analysers drift with temperature, humidity and time. Calibration with a syringe of known volume and a gas of known composition corrects them. If skipped, all results are off by the same percentage.

Is a portable metabolic system as accurate as a lab cart?

One validation found breath-by-breath mode ran about 6.6% below a stationary cart at the highest cycling load, and mixing-chamber mode about 5.8% above, while both were within small margins of the Douglas bag. Differences of a few percent between systems are ordinary.

What is a Douglas bag?

A Douglas bag is a large bag that collects exhaled air for a set time. The volume and gas fractions are measured afterwards. It has been the long-standing reference method for validating metabolic carts, though it is slow and gives no breath-by-breath data.

This article is general education. Train within your current ability and get coaching if you are unsure about technique.

References

  1. Validation of a portable metabolic system against the Douglas bag and a stationary cart PubMed Central
  2. Accuracy and reliability of the ParvoMedics TrueOne 2400 and MedGraphics VO2000 metabolic systems European Journal of Applied Physiology
  3. Validation of the Aerosport KB1-C portable metabolic system against the Douglas bag International Journal of Sports Medicine
Cibiraj Sathishkumar, sports scientist and founder of EvoQuest

Accredited Exercise and Sports Scientist, Founder of EvoQuest

Cibiraj studied Exercise and Sports Science at Deakin University in Australia and worked as a strength and conditioning coach for cricket and football clubs in Melbourne before founding EvoQuest in Coimbatore. He runs the performance assessments and writes every diagnostics report, so the person who measures you is the one who explains the numbers.

Full profile LinkedIn

Free 20-minute call

Want this applied to your sport and body?

Tell us the athlete's age, sport and goal. Cibiraj or the EvoQuest team will call or WhatsApp you, usually the same day. No obligation, no hard sell.

  • In person in Ramanathapuram, Coimbatore
  • Remote programmes for athletes abroad
  • Ages 9 and up, parents welcome on every call
Message us on WhatsApp

Or call +91 93440 48834