A dynamometer measures how hard a muscle pushes against something that does not move. Inside is a load cell: a metal block with strain gauges that change electrical resistance when it bends by a fraction of a millimetre. Electronics turn that tiny change into newtons, so the push becomes a number.

· Accredited Exercise and Sports Scientist, Founder of EvoQuest
Updated · 5 min read
Athlete pushing against a dynamometer during a strength test

Key takeaways

  • A strain gauge changes resistance when the metal under it stretches, and a bridge circuit turns that into a voltage.
  • Force becomes torque only when you multiply by the lever arm, so device position changes the result.
  • Hand-held devices are limited by the tester's own strength, not only the athlete's.
  • Peak force depends on how the signal is filtered and sampled.
  • EvoQuest uses dynamometry as one part of its lab assessment in Coimbatore.
On this page
  1. From push to number: the signal chain
  2. Force, torque and the lever arm
  3. Sampling, filtering and the peak
  4. Where the errors come from
  5. What published studies say about validity
  6. How EvoQuest uses dynamometry
  7. FAQs

From push to number: the signal chain

A dynamometer is a chain of four steps. The athlete pushes, a metal element deforms, an electronic circuit measures the deformation, and software converts it to force. Every dynamometer error sits in one of those four links.

The metal element is the load cell. Glued to it are strain gauges, thin foil grids whose electrical resistance rises slightly when the foil is stretched and falls when it is squeezed. The gauges are wired into a Wheatstone bridge, a circuit of four resistors that outputs a small voltage only when the resistances go out of balance. Push on the cell and the bridge voltage changes in proportion to the load. This is the same principle used inside bathroom scales and kitchen scales. For the plain-English version of what the device is, see what are dynamometers.

The voltage is small, often a few millivolts at full load, so an amplifier boosts it and an analogue-to-digital converter turns it into numbers. Software multiplies by a calibration factor that was found by hanging known weights on the cell. That factor is why a dynamometer needs a calibration check: if the factor drifts, every reading drifts with it.

Force, torque and the lever arm

Most strength tests ask about a joint, not a straight-line push, and joints rotate. A dynamometer measures force, but the muscle produces torque, which is force multiplied by the distance from the joint axis to where the force is applied.

A worked example shows the effect. Suppose a hand-held device is pressed on the shin and reads 300 N. If the pad sits 0.30 m below the knee joint, the knee extensor torque is 300 x 0.30 = 90 N·m. Move the pad to 0.25 m and the same torque reads 360 N. Two testers who place the pad at different heights get different force numbers from the same athlete.

That is why a good protocol records the lever arm, or fixes the pad position with a mark, and why comparing raw newtons across sessions is only fair when the position is identical. Fixed-frame devices, such as a strap-and-rig system for isometric tests, remove most of this by holding the position constant.

Sampling, filtering and the peak

A dynamometer that records force over time gives a curve, not one number. The curve rises, reaches a plateau and falls. The peak is the highest point on it, and where the peak lands depends on two choices made in software.

The first is sampling rate, the number of readings per second. Fast, short efforts need a higher rate than slow ones. If the rate is too low, the true peak can fall between two readings and be missed. The second is filtering. Raw signals carry electrical noise, so software smooths them. Heavy smoothing flattens the peak and delays it. Light smoothing leaves noise that can inflate the maximum. Two devices with different filters can give slightly different peaks for the same push.

Curve shape also gives information that a single peak hides. How fast the force rises in the first 100 to 200 milliseconds relates to explosive strength, and our post on rate of force development explains that measure. It only works if the sampling rate and filter are good enough to see the early rise.

Where the errors come from

Error in dynamometry comes from the device, the set-up and the people. The table lists the main sources and what reduces each.



The tester-strength problem deserves a note. In a hand-held "make" test the athlete pushes against the held device, and the tester must hold still. If the athlete is stronger than the tester, the limb moves and the reading reflects the tester's limit. Strong athletes and large muscle groups, such as the hip extensors, are the usual victims.

What published studies say about validity

Validity here means whether a cheaper or portable device agrees with a better reference. The usual reference for joint strength is the isokinetic machine, which is covered in how isokinetic testing works.

A systematic review by Stark and colleagues in PM&R compared hand-held dynamometry with isokinetic testing and concluded that hand-held devices are a reliable and valid option for clinical strength assessment, mainly because they are portable, easy to use and cheap. A later systematic review and meta-analysis by Chamorro and colleagues covered hip, knee and ankle strength. It rated the quality of the included studies as fair to poor on the COSMIN checklist and found that limits of agreement were narrower with isokinetic devices than with hand-held ones. Agreement between the two types varied by joint and movement.

The sensible rule is to compare an athlete only with their own earlier tests on the same device, tester and position. Our post on hand-held versus isokinetic dynamometers sets out when each makes sense.

How EvoQuest uses dynamometry

EvoQuest's three-hour assessment in Ramanathapuram, Coimbatore, includes dynamometry alongside force plates, timing gates and other measures. The coach fixes the position, the warm-up and the spoken cue so that a change in score reflects the athlete and not the set-up, and explains the result in plain language. The athlete-focused view is in dynamometers for athletes.

If you have pain or a recent injury, see a doctor or physio before strength testing. EvoQuest tests after clearance and does not diagnose or treat injuries.

If you want to know whether a test like this suits your sport or your child, start with a free 20-minute call on the call page or WhatsApp +91 93440 48834.

Frequently asked questions

How does a dynamometer measure force?

A dynamometer measures force with a load cell, a metal element carrying strain gauges. When the athlete pushes, the element flexes very slightly and the gauges change resistance. A bridge circuit converts that into a small voltage, and software multiplies it by a calibration factor to give newtons.

Is a hand-held dynamometer accurate?

It can be reliable when the protocol is fixed. A systematic review in PM&R judged hand-held dynamometry reliable and valid for clinical strength assessment. Accuracy drops when the tester cannot hold the device still against a strong athlete, so use the same tester and set-up each time.

Why does the position of the device change the reading?

Force and torque are different. Torque is force times the distance from the joint, so moving the pad changes the force reading even when the muscle effort is the same. Fixing and recording the pad position keeps results comparable between sessions.

Do I need to calibrate a dynamometer?

Yes. Strain gauge electronics can drift with time and temperature, so the device should be checked against a known mass on a regular schedule. If the check fails, readings should not be compared with older data until the device is corrected.

Can a dynamometer replace an isokinetic machine?

Not exactly. Hand-held devices are portable and cheap, but a meta-analysis found wider limits of agreement for them than isokinetic devices show with each other. Use one type consistently for tracking and avoid mixing the numbers.

This article is general education. Train within your current ability and get coaching if you are unsure about technique.

References

  1. Hand-held dynamometry correlation with the gold standard isokinetic dynamometry: a systematic review (Stark et al.) PM&R
  2. Reliability and validity of hand-held and isokinetic dynamometers for hip, knee and ankle strength: systematic review and meta-analysis (Chamorro et al.) Open Medicine
Cibiraj Sathishkumar, sports scientist and founder of EvoQuest

Accredited Exercise and Sports Scientist, Founder of EvoQuest

Cibiraj studied Exercise and Sports Science at Deakin University in Australia and worked as a strength and conditioning coach for cricket and football clubs in Melbourne before founding EvoQuest in Coimbatore. He runs the performance assessments and writes every diagnostics report, so the person who measures you is the one who explains the numbers.

Full profile LinkedIn

Free 20-minute call

Want this applied to your sport and body?

Tell us the athlete's age, sport and goal. Cibiraj or the EvoQuest team will call or WhatsApp you, usually the same day. No obligation, no hard sell.

  • In person in Ramanathapuram, Coimbatore
  • Remote programmes for athletes abroad
  • Ages 9 and up, parents welcome on every call
Message us on WhatsApp

Or call +91 93440 48834